Search facility:  
BIPM Home | Site map | Metrologia | KCDB | JCTLM-DB | Contact us  
   
BIPM policy on stating uncertainties
Version française
Traceability to the SI through the BIPM
Subject area:
What is traceability?
BIPM policy on stating uncertainties
Summary
BIPM services in the field of electricity
BIPM services in the field of ionizing radiation
BIPM calibrations and services in mass and related quantities
BIPM services in the field of time, frequency and gravimetry
Related articles
Calibration and characterization certificates issued by the BIPM
Ongoing comparisons conducted by the BIPM
KCDB
Direct access

The uncertainty associated with BIPM calibrations is a combined standard uncertainty, without the application of a coverage factor k.

This long-standing practice of not applying a coverage factor is considered to facilitate the combination of the BIPM and NMI uncertainties, and thus simplify the subsequent dissemination of the standard to the customers of the NMI. It can be assumed that the BIPM's measurements fulfil the criteria of section G6.6 of the GUM. In particular, for the purpose of calculating the expanded uncertainty for their end result at a specified level of confidence, an NMI can assume that the number of effective degrees of freedom for a BIPM calibration is sufficient to be able to use a coverage factor k = 2 for a level of confidence of approximately 95 %. Any exceptions are noted in the calibration certificate.

This policy was endorsed by the CIPM at its 93rd meeting (2004).



Related articles

Calibration and characterization certificates issued by the BIPM
Ongoing comparisons conducted by the BIPM